Blood Glucose Monitoring System2018-11-01T17:41:56+00:00

Blood Glucose Monitoring Systems List

TaiDoc has developed 5 generations of test strip with different level and fulfils the accuracy requirements of ISO 15197 : 2013 & EN ISO 15197 : 2015.

XPER, golden coating and laser engraving combine with 2+2 bio signal are the techniques to perform the stability, accuracy and measurement results.

XPER, golden coating and laser engraving combine with 2+2 bio signal are the techniques to perform the stability, accuracy and measurement results.

· Enzyme: GDH-FAD
· Sample size: 0.5 μL
· Reaction time: 5 seconds
· HCT range: 0 % – 70 %
· Glucose range: 10 – 800 mg/dL

TaiDoc patented the 2 + 2 (2 enzymes plus 2 signals) technology, which uses two different wells on the strips to detect HCT value by AC signal and glucose value by DC signal.

TaiDoc patented the 2 + 2 (2 enzymes plus 2 signals) technology, which uses two different wells on the strips to detect HCT value by AC signal and glucose value by DC signal.

· Enzyme: GDH-FAD
· Sample size: 1.0 μL
· Reaction time: 5 seconds
· HCT range: 0 % – 70 %
· Glucose range: 10 – 700 mg/dL

EF ( Electricity Fractionation ) technology utilize the negative charge characteristic on cell surface to separate blood cells from plasma for further estimation of hematocrit value.

1

Positive and negative charge will be created after applying blood sample.

Positive and negative charge will be created after applying blood sample.

Blood cells will be attracted to positive electrode.

Blood cells will be attracted to positive electrode.

The meter could simulate a hematocrit free condition from plasma phase and further estimate the HCT value to compensate the blood glucose value.

‧Enzyme: GDH-FAD
‧Sample size: 0.5 μL
‧Reaction time: 6 seconds
‧HCT range: 20 % – 65 %
‧Glucose range: 20 – 650 mg/dL

The meter could simulate a hematocrit free condition from plasma phase and further estimate the HCT value to compensate the blood glucose value.